Mythbusting: āUsing IQC and EQA From the Same Provider Leads to QC Bias’
Mythbusting: āUsing IQC and EQA From the Same Provider Leads to QC Bias’
Some laboratory professionals believe that using Internal Quality Control (IQC) and External Quality Assurance (EQA, also known as Proficiency Testing) material from the same provider can lead to increased levels of qc bias, or that their test system will not be appropriately challenged. It is important to address these concerns, because some labs may in fact be hindering their own performance by using IQC and EQA material from different sources.
It is important to first understand how IQC and EQA work together to help form a complete Laboratory Quality Management System.
IQC and EQA in Laboratory Quality Management
IQC is a means of monitoring test system precision on a daily basis. IQC effectively evaluates test system performance over time, so that any sudden or gradual shifts in performance can be detected. However, while IQC is an effective performance monitor, it cannot detect more intricate problems like calibration errors or wide acceptable limits provided by some QC manufacturers.
EQA is essential for challenging test system accuracy, and is carried out less frequently than IQC testing. EQA samples are tested ‘blind’ and the results are returned to the scheme organiser. As EQA testing compares an individual lab’s performance to other labs using the same method and instrument, it is a very effective tool for identification of potential issues.
Is there any disadvantage to using IQC and EQA material from the same provider?
The answer to this question depends primarily on the source material of the IQC and EQA. If an IQC provider manufactures their material using artificial additives or components of animal origin, then it will not be suitable to use EQA material from the same provider. Westgard (2011) maintains that using non-commutable IQC or EQA material can lead to results becoming compromised due to matrix effects – something which would not happen using commutable controls.
For example, with Immunoassay testing, non-human components of IQC material interact with antibodies in the reagent in a different way to fully human patient samples – ultimately giving unpredictable shifts, and not adhering to the ISO 15189 requirement to: “use quality control materials that react to the examining system in a manner as close as possible to patient samples”.
However, if the IQC and EQA material is manufactured using a source material which is similar in composition to patient samples (100% human), this commutable control will adequately mimic patient sample performance; meaning labs can use EQA and IQC material from the same provider with confidence that the integrity of their results is maintained.
Conclusion
ISO 15189 also states: “Use of independent third party control materials should be considered…”. In this instance, ‘Independent’ does not mean from a separate provider. It means that the QC material should not be optimized for use on one specific instrument (i.e. not dependent on a single instrument/method type).
No regulatory body states a requirement to use different providers for IQC and EQA material. Indeed, using IQC from one provider and EQA from another provider could increase the risk of labs using non-commutable material.
Labs should use commutable IQC and EQA material for a true assessment of their test system. Randox QC and RIQAS EQA are specifically designed with commutability in mind, giving labs a control which reflects patient sample performance and ensures excellent performance.
How can we help?
To learn how Randox can offer a complete solution for your laboratory, follow the links below or submit a question using the form above.
References
Westgard, S. (2011). Is QC Quality Compromised?. Available: https://www.westgard.com/qc-quality-compromised.htm. Last accessed 31st October 2017.